technical data
measuring ranges
0...2.000?s k=0.1, k=0.01
0...20.00?s k=1.0, k=0.1, k=0.01
0...200.0?s k=1.0, k=0.1, k=0.01
0...2.000ms k=1.0, k=0.1, inductive cell
0...20.00ms k=1.0, inductive cell
0...200.0ms inductive cell
0...2.000s inductive cell
range indicators
3 green lamps
display
red led-display 3?-digit, character high12.7mm
accuracy
0.5%
measuring frequency
80hz...7khz
measuring amplitude
70mv and 150mv, conductive cell only
step response
time between a conductivity change from 0% to 100% or reverse, measured between 10% and 90% = 4 seconds
temperature compensation
manual from 0 to 120?c, automatic by an external pt-100 platinum by 2 or 3-wire. the unit calculates with 25?c when the pt-100 sensor wires are broken
temperature slope
0.00%/?c (=without compensation) to 8.00%/?c.
conductivity of water
is measured and temperature compensated
reference temperature
25?c
maximum length of cell cable
cable capacity is compensated automatically. the max. capacity must be <0.02?f
limit or alarm contacts
two floating change-over contacts can be adjusted over the full range. each can be defined as a normally open or normally closed contact by an internal slide switch
adjustment of the limit contacts
each limit value can be adjusted by a button on the front panel. the limit value is displayed by pressing the corresponding button and can be changed with the + or - button over the full measuring range
status
status: the limit status is indicated by two green lamps
hysteresis
10 parts fixed
contacts rating
change over at 6a/230vac with resistive load
contact live
200`000 operations at max. load, 20`000`000 operation mechanically, without load
output current signal
0. . . 20ma or 4 . . . 20ma over the selected range. 0...20ma or 4...20ma can be defined by an internal slide switch which is accessible to the user. the output signal is galvanically isolated
output impedance
>1mw at 20ma
maximum load
>500w
supply voltage
230v/50-60hz, ca. 8va
ce-conformity
fulfilled
case
black plastic case 96 x 96 mm
mounting
2 mounting clamps
terminals
2 plug-in screw terminals for supply, limit contacts and signals
cut out
91 mm x 91 mm, depth 158 mm.
weight
670 grams (23.5 ounces)
options
different cell constants
leak proof housing ip54
supply voltages: 230vac
output signal voltage
limit contacts with different hysteresis
output current with signal isolation